BriefCase

Aidoc Medical, Ltd. • Radiology • 05/31/2019

510(k)IEC 62304 Class B
FDA Summary PDF

Device Purpose

The BriefCase device employs an artificial intelligence algorithm to analyze cervical spine CT images, identifying suspected linear lucencies compatible with fractures to facilitate expedited radiologist review. Its purpose is to improve workflow triage by providing rapid notifications that significantly reduce the time to radiologist attention for potentially critical findings, operating as an adjunct to standard diagnostic procedures without altering them.

Summary

The BriefCase device is a radiological computer-aided triage and notification software designed for analyzing cervical spine CT images. Its primary purpose is to assist hospital networks and trained radiologists in optimizing workflow triage by identifying and communicating suspected positive findings of linear lucencies indicative of cervical spine fractures. The system employs an artificial intelligence algorithm to process images and flag cases with detected findings. These notifications, which include compressed, non-diagnostic preview images, are presented on a standalone desktop application, running in parallel with standard image interpretation. The device does not alter original medical images and is not intended for diagnostic use, but rather to guide prioritization of image review based on professional judgment and other patient information. A multicenter, retrospective, blinded study involving 186 cases demonstrated the BriefCase software's performance. It achieved a sensitivity of 91.7% and a specificity of 88.6% in identifying CTs containing cervical spine fractures. Furthermore, the study evaluated the device's potential to reduce the time for radiologists to review critical scans. For true positive cases, BriefCase provided notifications in an average of 3.9 minutes, significantly faster than the standard-of-care mean time-to-exam-open of 58.4 minutes, suggesting an average potential time saving of 54.4 minutes. The device's Negative Predictive Value (NPV) was 99.0% and its Positive Predictive Value (PPV) was 47.2%, indicating its utility in accelerating the review of potentially critical cases.

AI Regulatory Concerns

Powered byJev by TypeSafe AI

Jev reads the full FDA letter and returns calibrated concern probabilities per regulatory category. These are AI estimates, not regulatory advice.

Keywords

Cervical Spine Fracture (CSF)Artificial Intelligence (AI)CT ImagesTriage SoftwareRadiology WorkflowNotification SystemImage AnalysisLinear LucenciesSensitivityTime-to-Notification

Device Metadata

Submission Number
K190896
Company
Aidoc Medical, Ltd.
Panel
Radiology
Product Code
N/A
Data Type
Images
Clinical Function
Assessment
AI Function
Analysis
AI Subclass
triage

Relevant Clinical Questions

  • Q1. How does the AI algorithm differentiate between true linear lucencies indicative of fractures and other artifacts or benign anatomical variants in cervical spine CTs?; What was the range of variability in 'time-to-exam-open' in the standard of care across the different clinical sites in the pivotal study, and how might this impact the generalizability of the reported time savings?; Given the Positive Predictive Value (PPV) of 47.2%, what is the anticipated impact on radiologist workload due to reviewing cases flagged by BriefCase that are ultimately negative for cervical spine fractures?; Were different types, locations, or severities of cervical spine fractures included in the study, and did the device's sensitivity vary significantly across these fracture characteristics?; How does the system handle variations in CT image quality or acquisition protocols, and what are the minimum image quality requirements for reliable algorithm performance?

Risk Classification

ESL Analysis
FDA Device ClassII

FDA Class II via 510(k) pathway; IEC 62304 Class B (software safety classification); Authorized before §524B enforcement

Learn how Parasoft automates IEC 62304 compliance testing →

Cybersecurity Evidence

Powered by SBOMator™

Based on AI-generated device summaries, not the full FDA submission. Items not detected here may still be present in the complete submission package.

Cyber Score10/100
SBOM Evidence⚠ Not mentioned in summary
Cyber Risk Assessment⚠ Not mentioned in summary
Postmarket Plan⚠ Not mentioned in summary
§524B ApplicableNo

Supply Chain BOM Evidence

Based on the CycloneDX / OWASP ECMA-424 xBOM ecosystem. Detection is from the AI-generated summary.

SBOMSoftware Bill of Materials
- Not detected
ML-BOMMachine Learning Bill of Materials(algorithm)
✓ Detected
SaaSBOMSoftware-as-a-Service Bill of Materials(network)
✓ Detected
HBOMHardware Bill of Materials
- Not detected
CBOMCryptography Bill of Materials
- Not detected
VDR/VEXVulnerability Disclosure Report / Vulnerability Exploitability eXchange
- Not detected
OBOMOperations Bill of Materials
- Not detected

Analysis:

• Analysis is based on the AI-generated device summary, not the full FDA submission package. Click "Analyze Full FDA PDF" below to scan the actual submission document.

• Evidence of ML-BOM, SaaSBOM detected in AI-generated device summary.

• This device was authorized before §524B took effect. Cybersecurity evidence requirements were less prescriptive at the time.

• Postmarket cybersecurity monitoring plan not clearly documented in AI-generated device summary.

• Not detected in AI-generated device summary: SBOM, HBOM, CBOM, VDR/VEX, OBOM. These may be present in the full FDA submission.