iSchemaView Inc. • Radiology • 07/09/2020

510(k)IEC 62304 Class B
FDA Summary PDF

Device Purpose

Rapid LVO 1.0 is designed to expedite the identification of potential Large Vessel Occlusions in CTA head scans, serving as a radiological computer-aided triage and notification tool. It employs a software algorithm to analyze images and alert clinicians to suspected LVOs, thereby aiming to improve workflow prioritization for patients with acute ischemic stroke without replacing standard diagnostic procedures.

Summary

Rapid LVO 1.0 is a radiological computer-aided triage and notification software specifically designed to analyze CTA head images for suspected Large Vessel Occlusions (LVOs). Its purpose is to assist hospital networks and trained radiologists by flagging potential LVO cases to streamline workflow triage. The system, a module within the broader Rapid Platform, utilizes a software algorithm to process DICOM-compliant CTA images, operating in parallel with standard image interpretation. It then issues notifications, including compressed preview images (for informational purposes only), to users via server, desktop application, email, or mobile, highlighting cases that may require urgent attention for acute ischemic stroke. The software does not modify original medical images and is not a standalone diagnostic tool; its outputs are intended to support clinical judgment and image prioritization. Performance validation demonstrated high accuracy with a sensitivity of 0.970 and specificity of 0.956 for LVO detection, and an ROC AUC of 0.99. A key operational feature is its rapid notification capability, achieving an average time-to-notification of 2.86 minutes from image acquisition to alert delivery. Clinicians are always responsible for reviewing full, non-compressed images on a diagnostic viewer before making patient care decisions.

AI Regulatory Concerns

Powered byJev by TypeSafe AI

Jev reads the full FDA letter and returns calibrated concern probabilities per regulatory category. These are AI estimates, not regulatory advice.

Keywords

Large Vessel Occlusion (LVO)CTA Head ImagesTriage SoftwareNotification SystemRadiological AIWorkflow PrioritizationDICOMImage Analysis AlgorithmAcute Ischemic StrokeSensitivity/Specificity

Device Metadata

Submission Number
K200941
Company
iSchemaView Inc.
Panel
Radiology
Product Code
N/A
Data Type
Images
Clinical Function
Assessment
AI Function
Analysis
AI Subclass
triage

Relevant Clinical Questions

  • Q1. How does the Rapid LVO algorithm differentiate between true LVOs and mimics, such as severe stenosis or artifacts, in CTA imaging?; What impact does the 2.86-minute notification time have on 'door-to-treatment' times in real-world clinical settings for LVO patients?; Given the performance metrics (Se 0.970, Sp 0.956), what is the anticipated rate of false positives and false negatives in a typical stroke center population, and what are the clinical implications?; How does the system handle variations in CTA image quality, contrast timing, or scanner protocols, and does this impact its detection accuracy?; Beyond simple notification, are there plans to integrate advanced quantitative LVO characteristics (e.g., clot burden, location detail) into the preview information provided by Rapid LVO?

Risk Classification

ESL Analysis
FDA Device ClassII

FDA Class II via 510(k) pathway; IEC 62304 Class B (software safety classification); Authorized before §524B enforcement

Learn how Parasoft automates IEC 62304 compliance testing →

Cybersecurity Evidence

Powered by SBOMator™

Based on AI-generated device summaries, not the full FDA submission. Items not detected here may still be present in the complete submission package.

Cyber Score30/100
SBOM Evidence⚠ Not mentioned in summary
Cyber Risk Assessment⚠ Not mentioned in summary
Postmarket Plan✓ Detected in Summary
§524B ApplicableNo

Supply Chain BOM Evidence

Based on the CycloneDX / OWASP ECMA-424 xBOM ecosystem. Detection is from the AI-generated summary.

SBOMSoftware Bill of Materials
- Not detected
ML-BOMMachine Learning Bill of Materials(algorithm)
✓ Detected
SaaSBOMSoftware-as-a-Service Bill of Materials(server, network, platform)
✓ Detected
HBOMHardware Bill of Materials
- Not detected
CBOMCryptography Bill of Materials
- Not detected
VDR/VEXVulnerability Disclosure Report / Vulnerability Exploitability eXchange
- Not detected
OBOMOperations Bill of Materials(operational)
✓ Detected

Analysis:

• Analysis is based on the AI-generated device summary, not the full FDA submission package. Click "Analyze Full FDA PDF" below to scan the actual submission document.

• Evidence of ML-BOM, SaaSBOM, OBOM detected in AI-generated device summary.

• This device was authorized before §524B took effect. Cybersecurity evidence requirements were less prescriptive at the time.

• Not detected in AI-generated device summary: SBOM, HBOM, CBOM, VDR/VEX. These may be present in the full FDA submission.

Preparing a similar submission?

ESL takes complete ownership of your software evidence work - SBOM & CVE remediation with SBOMator™, static analysis & unit testing with Parasoft, traceability, and full IEC 62304 lifecycle support.

Assess Your Readiness →Or contact ESL directly
Rapid LVO - iSchemaView Inc. | ESL FDA AI Device Intelligence