ClariSIGMAM

ClariPi Inc. • Radiology • 09/10/2021

510(k)IEC 62304 Class B
FDA Summary PDF

Device Purpose

ClariSIGMAM is an automated image processing software designed to enhance breast cancer screening by objectively quantifying breast density from 2D digital mammograms. It utilizes algorithms to calculate the ratio of fibroglandular tissue to total breast area, providing a binary BI-RADS density classification (fatty vs. dense) as adjunctive information for clinicians.

Summary

ClariSIGMAM is a standalone software application designed to automatically analyze "for presentation" 2D digital mammograms to aid in assessing breast tissue composition. It functions by calculating the percent breast density, defined as the ratio of fibroglandular tissue area to total breast area, for each breast. The software then categorizes breast density into two groups: BI-RADS A+B (fatty) and BI-RADS C+D (dense), aligning with the American College of Radiology’s BI-RADS density classification. This information, including specific area measurements (cm²) for breast and fibroglandular tissue and percent density, is outputted as a DICOM structured report or secondary capture for display on mammography workstations or PACS. ClariSIGMAM is intended for use with compatible Hologic and GE full-field digital mammography systems and provides adjunctive information, not serving as a direct diagnostic aid. The performance of ClariSIGMAM was validated through non-clinical testing, comparing its automated breast density estimates against a gold standard established by expert radiologists using interactive thresholding software (Cumulus). Validation also assessed reproducibility concerning age-related density decrease, consistency over time with paired mammograms, and similarity in density estimates between left and right breasts. In a key comparison with a consensus visual assessment by four expert readers on a dataset of 837 mammograms, ClariSIGMAM demonstrated substantial agreement for binary breast density classification (fatty vs. dense), achieving a Kappa statistic of 0.734. The software showed a sensitivity of 86.6% in identifying consensus-defined fatty breasts and a sensitivity of 87.3% in identifying consensus-defined dense breasts. The software meets established international standards and guidance for medical device software and interoperability.

AI Regulatory Concerns

Powered byJev by TypeSafe AI

Jev reads the full FDA letter and returns calibrated concern probabilities per regulatory category. These are AI estimates, not regulatory advice.

Keywords

Breast DensityMammographyImage Processing SoftwareFibroglandular TissueBI-RADSDICOMAutomated AssessmentAdjunctive InformationPercent Breast DensityReproducibility

Device Metadata

Submission Number
K203785
Company
ClariPi Inc.
Panel
Radiology
Product Code
N/A
Data Type
Images
Clinical Function
Assessment
AI Function
Analysis
AI Subclass
quantification/feature localization

Relevant Clinical Questions

  • Q1. How does ClariSIGMAM's algorithm differentiate between fibroglandular tissue and other breast tissues in 2D mammograms, and what are its limitations with overlapping tissues?; Given the Kappa value of 0.734, what were the primary sources of disagreement between ClariSIGMAM's automated assessment and the expert reader consensus for binary density classification?; The software outputs a binary classification (BI-RADS A+B vs. C+D). How does this simplification compare to the clinical utility of the full four-category BI-RADS assessment, particularly for borderline cases between B and C?; What specific image quality parameters or acquisition artifacts from the compatible Hologic and GE systems most significantly impact the accuracy of ClariSIGMAM's density calculations?; How were the 'substantial data sets' for validation testing characterized in terms of patient demographics (age, ethnicity), BI-RADS density distribution, and image acquisition variability, and how might these factors influence real-world performance?

Risk Classification

ESL Analysis
FDA Device ClassII

FDA Class II via 510(k) pathway; IEC 62304 Class B (software safety classification); Authorized before §524B enforcement

Learn how Parasoft automates IEC 62304 compliance testing →

Cybersecurity Evidence

Powered by SBOMator™

Based on AI-generated device summaries, not the full FDA submission. Items not detected here may still be present in the complete submission package.

Cyber Score25/100
SBOM Evidence⚠ Not mentioned in summary
Cyber Risk Assessment⚠ Not mentioned in summary
Postmarket Plan✓ Detected in Summary
§524B ApplicableNo

Supply Chain BOM Evidence

Based on the CycloneDX / OWASP ECMA-424 xBOM ecosystem. Detection is from the AI-generated summary.

SBOMSoftware Bill of Materials
- Not detected
ML-BOMMachine Learning Bill of Materials(algorithm, dataset, data set)
✓ Detected
SaaSBOMSoftware-as-a-Service Bill of Materials(server)
✓ Detected
HBOMHardware Bill of Materials
- Not detected
CBOMCryptography Bill of Materials
- Not detected
VDR/VEXVulnerability Disclosure Report / Vulnerability Exploitability eXchange
- Not detected
OBOMOperations Bill of Materials
- Not detected

Analysis:

• Analysis is based on the AI-generated device summary, not the full FDA submission package. Click "Analyze Full FDA PDF" below to scan the actual submission document.

• Evidence of ML-BOM, SaaSBOM detected in AI-generated device summary.

• This device was authorized before §524B took effect. Cybersecurity evidence requirements were less prescriptive at the time.

• Not detected in AI-generated device summary: SBOM, HBOM, CBOM, VDR/VEX, OBOM. These may be present in the full FDA submission.

Preparing a similar submission?

ESL takes complete ownership of your software evidence work - SBOM & CVE remediation with SBOMator™, static analysis & unit testing with Parasoft, traceability, and full IEC 62304 lifecycle support.

Assess Your Readiness →Or contact ESL directly
ClariSIGMAM - ClariPi Inc. | ESL FDA AI Device Intelligence