Medscape Medical NewsAll articles

Formus Hip

Formus Labs, Ltd • Radiology • 03/31/2023

510(k)IEC 62304 Class B
FDA Summary PDF

Device Purpose

Formus Hip is a Software as a Medical Device that facilitates pre-operative planning for total hip arthroplasty by using AI-driven image segmentation and statistical shape modeling from CT scans to generate patient-specific 3D anatomical models and propose implant configurations. Non-clinical validation demonstrated high accuracy in 3D model generation compared to manual segmentation and strong concordance with surgeon-selected implant sizes.

Summary

Formus Hip is a semi-automated Software as a Medical Device (SaMD) designed for pre-operative planning of primary total hip arthroplasty, specifically utilizing the Zimmer Biomet Taperloc G7 system. The software imports patient pre-dimensioned CT scans and employs AI-based automatic image segmentation algorithms, trained on diverse patient data (ages 21-94), to create a 3D model of the patient's femur and pelvis without initial user input. It further uses statistical shape models, trained on subjects aged 18-89, and algorithms to fit the femoral stem and acetabular cup based on the patient's anatomy. The system includes an integrated database of orthopedic hip implant geometries for overlay and allows healthcare professionals to interactively customize the surgical plan via a graphical user interface, with finalized plans exportable as PDF reports. The device's non-clinical performance was validated for image processing accuracy and implant sizing. Image processing accuracy was assessed using 60 US patient CT scans, comparing Formus Hip's 3D models to manual segmentations by radiologists. Results met acceptance criteria, with average Sorensen-Dice scores of 0.95 (hemipelvis) and 0.97 (femur), and mean absolute distances (MAD) below 2mm. Statistical shape modeling accuracy also met targets. Implant sizing validation, conducted on 133 Australian patient images, compared software-recommended sizes to those chosen by orthopedic surgeons using traditional methods. The software achieved 94% agreement for stem sizes and 98% for cup sizes within ±2 sizes of the surgeons' choices, meeting the performance goal of at least 80%.

AI Regulatory Concerns

Powered byJev by TypeSafe AI

Jev reads the full FDA letter and returns calibrated concern probabilities per regulatory category. These are AI estimates, not regulatory advice.

Keywords

Pre-operative planningOrthopedic surgeryTotal hip arthroplastyCT scansAI-based segmentation3D modelingStatistical shape modelsImplant sizingImage processing accuracySaMD (Software as a Medical Device)

Device Metadata

Submission Number
K213272
Company
Formus Labs, Ltd
Panel
Radiology
Product Code
N/A
Data Type
Images
Clinical Function
Intervention
AI Function
Analysis
AI Subclass
quantification/feature localization

Relevant Clinical Questions

  • Q1. How does the AI-based automatic image segmentation algorithm handle CT scans with significant artifacts or unusual anatomical variations not well-represented in the training dataset (ages 21-94 for segmentation, 18-89 for shape models)?; What is the specific performance of the statistical shape models in cases of extreme anatomical deformities, such as severe dysplasia or post-traumatic changes, compared to typical anatomy within the training data parameters?; Given that implant sizing validation was performed on Australian patient data, what specific measures ensure its generalizability and accuracy across diverse global populations and varying surgical preferences?; What is the typical processing time for Formus Hip to generate a complete pre-operative plan from a patient's CT scan, and how does this integrate into existing clinical workflows?; Beyond the Zimmer Biomet Taperloc G7 system, what is the protocol and anticipated timeline for incorporating and validating additional hip implant systems from other manufacturers into the software's database?

Risk Classification

ESL Analysis
FDA Device ClassII

FDA Class II via 510(k) pathway; IEC 62304 Class B (software safety classification); Authorized before §524B enforcement

Learn how Parasoft automates IEC 62304 compliance testing →

Cybersecurity Evidence

Powered by SBOMator™

Based on AI-generated device summaries, not the full FDA submission. Items not detected here may still be present in the complete submission package.

Cyber Score10/100
SBOM Evidence⚠ Not mentioned in summary
Cyber Risk Assessment⚠ Not mentioned in summary
Postmarket Plan⚠ Not mentioned in summary
§524B ApplicableNo

Supply Chain BOM Evidence

Based on the CycloneDX / OWASP ECMA-424 xBOM ecosystem. Detection is from the AI-generated summary.

SBOMSoftware Bill of Materials
- Not detected
ML-BOMMachine Learning Bill of Materials(training data, algorithm, dataset)
✓ Detected
SaaSBOMSoftware-as-a-Service Bill of Materials
- Not detected
HBOMHardware Bill of Materials
- Not detected
CBOMCryptography Bill of Materials
- Not detected
VDR/VEXVulnerability Disclosure Report / Vulnerability Exploitability eXchange
- Not detected
OBOMOperations Bill of Materials(configuration)
✓ Detected

Analysis:

• Analysis is based on the AI-generated device summary, not the full FDA submission package. Click "Analyze Full FDA PDF" below to scan the actual submission document.

• Evidence of ML-BOM, OBOM detected in AI-generated device summary.

• This device was authorized before §524B took effect. Cybersecurity evidence requirements were less prescriptive at the time.

• Postmarket cybersecurity monitoring plan not clearly documented in AI-generated device summary.

• Not detected in AI-generated device summary: SBOM, SaaSBOM, HBOM, CBOM, VDR/VEX. These may be present in the full FDA submission.

Preparing a similar submission?

ESL takes complete ownership of your software evidence work - SBOM & CVE remediation with SBOMator™, static analysis & unit testing with Parasoft, traceability, and full IEC 62304 lifecycle support.

Assess Your Readiness →Or contact ESL directly
Formus Hip - Formus Labs, Ltd | ESL FDA AI Device Intelligence