Medscape Medical NewsAll articles

autoSCORE

Holberg EEG AS • Neurology • 01/07/2024

510(k)IEC 62304 Class B§524B Cyber Device
FDA Summary PDF

Device Purpose

autoSCORE is a software-only device utilizing a deep learning Convolutional Neural Network to assist clinicians by analyzing scalp EEG recordings for both epileptiform and non-epileptiform abnormalities, providing probabilistic assessments and marker suggestions. Its methodology aims to enhance the efficiency and consistency of EEG review, with clinical studies demonstrating its ability to identify abnormalities with varying degrees of accuracy compared to human experts and predicate technologies.

Summary

autoSCORE is a software-only decision support tool intended to assist qualified medical practitioners in reviewing, monitoring, and analyzing scalp EEG recordings from patients aged three months and older, excluding neonatal and ICU settings. It employs a deep learning algorithm, specifically a Convolutional Neural Network trained on a large dataset, to process previously acquired EEG data. The software identifies and marks potential spikes corresponding to epileptiform abnormalities (Focal and Generalized) and also detects and classifies non-epileptiform abnormalities (Focal and Diffuse Non-epileptiform). It assesses the probability of these abnormalities and sends this information, along with suggested marker placements, to compatible EEG review software, without altering the original EEG data or providing a diagnostic conclusion. Clinical validation across five retrospective studies demonstrated autoSCORE's capabilities. When classifying EEGs as normal or abnormal at the recording level, it achieved high agreement with human experts (HEs), with sensitivities often above 80% and specificities generally between 84% and 96%. For specific abnormality types like Focal Epileptiform and Generalized Epileptiform, performance varied, with notable strengths in negative predictive value and correlation with HE assessments; for instance, Generalized Epileptiform detection showed sensitivities from 71.2% to 100%. While marker-level agreement was more modest, it improved with higher autoSCORE-assigned probability, and in comparisons for epileptiform detection, autoSCORE generally outperformed predicate devices in accuracy and specificity.

AI Regulatory Concerns

Powered byJev by TypeSafe AI

Jev reads the full FDA letter and returns calibrated concern probabilities per regulatory category. These are AI estimates, not regulatory advice.

Keywords

EEG analysisDecision support softwareDeep learningConvolutional Neural NetworkSpike detectionEpileptiform abnormalitiesNon-epileptiform abnormalitiesAbnormality probabilityClinical validationScalp EEG

Device Metadata

Submission Number
K231068
Company
Holberg EEG AS
Panel
Neurology
Product Code
N/A
Data Type
Signals
Clinical Function
Assessment
AI Function
Analysis
AI Subclass
detection/diagnosis

Relevant Clinical Questions

  • Q1. How does the performance of autoSCORE's non-epileptiform abnormality detection compare, in terms of specific metrics (sensitivity, specificity, PPV, NPV), to established methods or expert consensus for these less commonly automated findings?; What is the variability in autoSCORE's performance across different EEG recording montages, durations, or signal quality levels commonly encountered in routine clinical practice?; Given that marker-level agreement with human experts is lower than recording-level agreement, what are the primary types of discrepancies (e.g., over-marking, under-marking, misclassification of marker type), and how might these impact clinical workflow efficiency?; The validation studies used human expert assessments as a reference; how does inter-rater reliability among the human experts themselves compare to the agreement levels observed between autoSCORE and the expert consensus, particularly for nuanced or rare abnormalities?; How does the 'probability' output for abnormalities correlate with specific electrographic features or clinical contexts, and what is the optimal threshold for these probabilities to balance sensitivity and specificity for different clinical decision-making scenarios?

Risk Classification

ESL Analysis
FDA Device ClassII

FDA Class II via 510(k) pathway; IEC 62304 Class B (software safety classification); Qualifies as a cyber device under §524B (post-Dec 2023, networked)

Learn how Parasoft automates IEC 62304 compliance testing →

Cybersecurity Evidence

Powered by SBOMator™

Based on AI-generated device summaries, not the full FDA submission. Items not detected here may still be present in the complete submission package.

Cyber Score25/100
SBOM Evidence⚠ Not in summary - likely in full submission per §524B
Cyber Risk Assessment⚠ Not in summary - likely in full submission per §524B
Postmarket Plan✓ Detected in Summary
§524B ApplicableYes

Supply Chain BOM Evidence

Based on the CycloneDX / OWASP ECMA-424 xBOM ecosystem. Detection is from the AI-generated summary.

SBOMSoftware Bill of Materials
- Not detected
ML-BOMMachine Learning Bill of Materials(deep learning, neural network, algorithm, dataset)
✓ Detected
SaaSBOMSoftware-as-a-Service Bill of Materials(network)
✓ Detected
HBOMHardware Bill of Materials
- Not detected
CBOMCryptography Bill of Materials
- Not detected
VDR/VEXVulnerability Disclosure Report / Vulnerability Exploitability eXchange
- Not detected
OBOMOperations Bill of Materials
- Not detected

Analysis:

• Analysis is based on the AI-generated device summary, not the full FDA submission package. Click "Analyze Full FDA PDF" below to scan the actual submission document.

• Evidence of ML-BOM, SaaSBOM detected in AI-generated device summary.

• SBOM not mentioned in AI-generated device summary. Under §524B (effective Oct 2023), cyber devices must provide a SBOM.

• Cybersecurity risk assessment not referenced in AI-generated device summary. FDA cybersecurity guidance (2023) expects this for cyber devices.

• Not detected in AI-generated device summary: SBOM, HBOM, CBOM, VDR/VEX, OBOM. These may be present in the full FDA submission.

Preparing a similar submission?

ESL takes complete ownership of your software evidence work - SBOM & CVE remediation with SBOMator™, static analysis & unit testing with Parasoft, traceability, and full IEC 62304 lifecycle support.

Assess Your Readiness →Or contact ESL directly
autoSCORE - Holberg EEG AS | ESL FDA AI Device Intelligence