Medscape Medical NewsAll articles

Tyto Insights for Wheeze Detection

Tyto Care Ltd. • Anesthesiology • 12/13/2023

510(k)IEC 62304 Class C
FDA Summary PDF

Device Purpose

The Tyto Insights for Wheeze Detection is an AI-driven software system that analyzes lung sound recordings from a compatible Tyto Stethoscope to assist in identifying suspected wheezing in adults and pediatric patients (2 years and older). Its core science involves an artificial intelligence algorithm, trained on real-world acoustic data, which processes these sounds to provide a binary output (wheeze suspected/not suspected); validation studies showed a sensitivity of 0.54, specificity of 0.98, and non-inferiority to a rule-based predicate device with a higher AUC.

Summary

The Tyto Insights for Wheeze Detection is an over-the-counter, artificial intelligence-enabled software system designed for evaluating lung sounds in adults and children aged two years and older. It functions by automatically analyzing acoustic signals of the lungs, as recorded by a compatible Tyto Stethoscope, to identify instances where wheezing is suspected. The system comprises an Application Server, an AI-driven Algorithm Server that processes audio files and outputs a wheeze detection result (positive, negative, or unable to analyze), and a Web Server that graphically displays this result to both patients and clinicians. It is explicitly intended only for wheeze detection and not for identifying other lung sounds, with the results requiring interpretation by a licensed healthcare professional in conjunction with other relevant patient information. The device's performance was evaluated retrospectively using a dataset of 371 lung sound recordings from 359 patients (ages 2 and older), with ground truth established by a majority vote of three blinded pulmonologists. For stand-alone accuracy, this evaluation yielded a sensitivity of 0.54 and a specificity of 0.98 for wheeze detection. In a comparative analysis against its predicate device, which uses a rule-based algorithm, the Tyto Insights for Wheeze Detection met its primary endpoint of non-inferiority and demonstrated an Area Under the Curve (AUC) of 0.9654, numerically higher than the predicate's AUC of 0.9085, for detecting wheezes. The AI algorithm was trained using real-world recordings from the compatible Tyto Stethoscope, and the system processes recordings in segments up to 12 seconds, not processing signals shorter than 6 seconds.

AI Regulatory Concerns

Powered byJev by TypeSafe AI

Jev reads the full FDA letter and returns calibrated concern probabilities per regulatory category. These are AI estimates, not regulatory advice.

Keywords

Wheeze DetectionArtificial Intelligence (AI)Lung SoundsAcoustic AnalysisDecision Support SoftwareTyto StethoscopePediatricsOver-The-Counter (OTC)SensitivitySpecificity

Device Metadata

Submission Number
K232237
Company
Tyto Care Ltd.
Panel
Anesthesiology
Product Code
N/A
Data Type
Signals
Clinical Function
Assessment
AI Function
Analysis
AI Subclass
diagnosis

Relevant Clinical Questions

  • Q1. Given the sensitivity of 0.54, what is the potential clinical impact of false negatives, especially in acute respiratory distress scenarios, and how should users be guided to mitigate this risk?; How does the AI algorithm differentiate wheezes from other adventitious lung sounds (e.g., rhonchi, crackles) or common background noises, and what were the specific features or datasets used for training this differentiation?; What is the variability in performance (sensitivity/specificity) across different age sub-groups within the pediatric population (e.g., 2-5 years vs. 6-12 years vs. adolescents) and in patients with varying severities of wheezing or comorbidities other than COPD or Asthma?; How does the device's performance change in real-world scenarios with varying levels of user proficiency in stethoscope placement and recording quality, particularly for lay users in a home environment?; Considering the Positive Predictive Value of 0.72, what further diagnostic steps or clinical assessments are typically recommended following a 'wheeze suspected' result from the device to confirm the diagnosis and determine the underlying etiology?

Risk Classification

ESL Analysis
FDA Device ClassII

FDA Class II via 510(k) pathway; IEC 62304 Class C (software safety classification); Post-§524B but no clear network capability detected

Learn how Parasoft automates IEC 62304 compliance testing →

Cybersecurity Evidence

Powered by SBOMator™

Based on AI-generated device summaries, not the full FDA submission. Items not detected here may still be present in the complete submission package.

Cyber Score30/100
SBOM Evidence⚠ Not in summary - likely in full submission per §524B
Cyber Risk Assessment⚠ Not in summary - likely in full submission per §524B
Postmarket Plan✓ Detected in Summary
§524B ApplicableYes

Supply Chain BOM Evidence

Based on the CycloneDX / OWASP ECMA-424 xBOM ecosystem. Detection is from the AI-generated summary.

SBOMSoftware Bill of Materials
- Not detected
ML-BOMMachine Learning Bill of Materials(algorithm, dataset, ground truth)
✓ Detected
SaaSBOMSoftware-as-a-Service Bill of Materials(server, remote)
✓ Detected
HBOMHardware Bill of Materials
- Not detected
CBOMCryptography Bill of Materials
- Not detected
VDR/VEXVulnerability Disclosure Report / Vulnerability Exploitability eXchange
- Not detected
OBOMOperations Bill of Materials(environment)
✓ Detected

Analysis:

• Analysis is based on the AI-generated device summary, not the full FDA submission package. Click "Analyze Full FDA PDF" below to scan the actual submission document.

• Evidence of ML-BOM, SaaSBOM, OBOM detected in AI-generated device summary.

• SBOM not mentioned in AI-generated device summary. Under §524B (effective Oct 2023), cyber devices must provide a SBOM.

• Cybersecurity risk assessment not referenced in AI-generated device summary. FDA cybersecurity guidance (2023) expects this for cyber devices.

• Not detected in AI-generated device summary: SBOM, HBOM, CBOM, VDR/VEX. These may be present in the full FDA submission.

Preparing a similar submission?

ESL takes complete ownership of your software evidence work - SBOM & CVE remediation with SBOMator™, static analysis & unit testing with Parasoft, traceability, and full IEC 62304 lifecycle support.

Assess Your Readiness →Or contact ESL directly
Tyto Insights for Wheeze Detection - Tyto Care Ltd. | ESL FDA AI Device Intelligence