Tempus ECG-AF
Tempus AI, Inc. • Cardiovascular • 06/21/2024
Device Purpose
Tempus ECG-AF is a machine learning-based software that analyzes 12-lead ECG recordings from specific asymptomatic patients aged 65 and older to identify signs indicative of an increased risk of developing atrial fibrillation or atrial flutter within the next 12 months. The technology processes standard ECG data, patient age, and sex through a trained algorithm to provide a risk notification, intended to support clinical decisions for further diagnostic follow-up rather than for direct diagnosis or treatment guidance.
Summary
The Tempus ECG-AF device is a software tool utilizing machine learning to analyze 12-lead resting electrocardiogram (ECG) recordings. It is intended for use in patients aged 65 years and older who do not have a pre-existing diagnosis of atrial fibrillation (AF) or atrial flutter, are without pacemakers or implantable cardioverter defibrillators, and have not had cardiac surgery within the preceding 30 days. The software's purpose is to detect subtle signs within these ECGs that are associated with an increased likelihood of the patient experiencing AF or atrial flutter within the next 12 months, thereby aiding healthcare providers in identifying individuals who may warrant further diagnostic evaluation for potential undiagnosed AF. The core of Tempus ECG-AF is a 'locked' machine-learning model, trained on a large dataset exceeding 1.5 million ECGs, which processes the patient's 12-lead ECG tracing along with their age and sex. This analysis generates an uncalibrated risk score, subsequently converted into a binary output—either 'increased risk' or 'no increased risk'—based on pre-defined thresholds, or an 'unclassifiable' result if data quality is insufficient. Clinical validation in a retrospective study of 4017 patients demonstrated a sensitivity of 31% and a specificity of 92% for identifying patients who would be diagnosed with AF within 12 months. The positive predictive value (PPV) was 19% and the negative predictive value (NPV) was 95% in a study population with a 6.0% one-year AF incidence, suggesting that a positive result identifies a patient group with a roughly threefold higher likelihood of developing AF compared to the baseline.
AI Regulatory Concerns
Powered by
Jev reads the full FDA letter and returns calibrated concern probabilities per regulatory category. These are AI estimates, not regulatory advice.
Keywords
Device Metadata
Relevant Clinical Questions
- Q1. Given the sensitivity of 31%, what proportion of patients who will develop AF within 12 months would be missed by this screening tool, and what are the clinical implications of these false negatives?; How does the model's performance vary across different ECG machine manufacturers and models beyond the specified GE and Philips systems, and with different electrode types or sampling rates not explicitly validated?; What specific ECG features or subtle patterns does the machine learning algorithm primarily rely on to identify an increased risk of future AF, and are these features novel or extensions of known subclinical ECG markers?; Considering the training dataset was predominantly White (96%), and the validation set was 82% White, what is the anticipated performance in more racially diverse populations, and are there plans for specific validation in these groups?; Since repeated testing performance has not been evaluated, what is the recommended clinical approach or testing interval if a patient initially receives a 'no increased risk' result but remains in a high-risk demographic for AF?
Risk Classification
ESL AnalysisFDA Class II via 510(k) pathway; IEC 62304 Class C (software safety classification); Post-§524B but no clear network capability detected
Learn how Parasoft automates IEC 62304 compliance testing →Cybersecurity Evidence
Powered by SBOMator™Based on AI-generated device summaries, not the full FDA submission. Items not detected here may still be present in the complete submission package.
Supply Chain BOM Evidence
Based on the CycloneDX / OWASP ECMA-424 xBOM ecosystem. Detection is from the AI-generated summary.
Analysis:
• Analysis is based on the AI-generated device summary, not the full FDA submission package. Click "Analyze Full FDA PDF" below to scan the actual submission document.
• Evidence of ML-BOM detected in AI-generated device summary.
• SBOM not mentioned in AI-generated device summary. Under §524B (effective Oct 2023), cyber devices must provide a SBOM.
• Cybersecurity risk assessment not referenced in AI-generated device summary. FDA cybersecurity guidance (2023) expects this for cyber devices.
• Postmarket cybersecurity monitoring plan not clearly documented in AI-generated device summary.
• Not detected in AI-generated device summary: SBOM, SaaSBOM, HBOM, CBOM, VDR/VEX, OBOM. These may be present in the full FDA submission.