AEYE-DS
AEYE Health Inc. • Ophthalmic • 04/23/2024
Device Purpose
The AEYE-DS is an artificial intelligence-based software designed for healthcare providers to automatically analyze retinal images from compatible fundus cameras (Topcon NW400, Optomed Aurora) for the detection of more than mild diabetic retinopathy in adult diabetic patients. Clinical studies have demonstrated its diagnostic performance, showing high sensitivity, specificity, and imageability, supporting its role as a screening tool for this condition.
Summary
The AEYE-DS is a retinal diagnostic software device utilizing artificial intelligence to automatically detect more than mild diabetic retinopathy (mtmDR) in adults diagnosed with diabetes who have not previously been diagnosed with diabetic retinopathy. It is intended for use by healthcare providers and functions by analyzing fundus retinal images acquired using compatible cameras, specifically the Topcon NW400 and the Optomed Aurora. The system architecture comprises five software components: Client (local user interface), Service, Analytics, Reporting and Archiving, and System Security, with the latter components typically server-based and accessed via a secure internet connection. The Client module guides image acquisition and transfer; the Service module manages communication and image transfer to the Analytics module, which processes images and returns diagnostic information regarding image quality and the presence or absence of mtmDR. Clinical performance for the AEYE-DS with the Optomed Aurora camera was demonstrated in two prospective, multi-center studies. In Study 1 (317 subjects), fundus-based sensitivity was 92% and specificity was 94%, with a Positive Predictive Value (PPV) of 68% and Negative Predictive Value (NPV) of 99%. In Study 2 (362 subjects), fundus-based sensitivity was 93% and specificity was 89%, with a PPV of 53% and NPV of 99%. Imageability was 99% in both studies. The device successfully identified all participants with ETDRS level 43 or higher in both studies. A precision study demonstrated high intra-operator repeatability (99%), between-operator reproducibility (95%), and between-device reproducibility (97%) with the Optomed Aurora camera. A human factors validation study confirmed ease of use and user manual comprehension by potential users after initial training.
AI Regulatory Concerns
Powered by
Jev reads the full FDA letter and returns calibrated concern probabilities per regulatory category. These are AI estimates, not regulatory advice.
Keywords
Device Metadata
Relevant Clinical Questions
- Q1. How does the AI algorithm differentiate between 'more than mild diabetic retinopathy' and milder forms or other retinal pathologies visible in fundus images?; What are the specific image quality parameters the AEYE-DS software uses to determine if an image is 'insufficient quality,' and how might this impact the workflow and rescan rates in a busy clinical setting?; Given the variation in Positive Predictive Value (PPV) between Study 1 (68%) and Study 2 (53%) using the Optomed Aurora camera, what are the potential contributing factors to this difference, considering the study populations were not enriched?; The document mentions that 'novice operators' obtained fundoscopy images. What level of training was provided to these operators, and how does this compare to the expected training for typical healthcare providers intended to use the AEYE-DS system?; Study 2 indicated a significant effect of age on performance, with the lower age group (<55 years) showing better results. What are the hypothesized reasons for this age-related performance difference, and are there specific considerations or adjustments recommended when using the device with older patient populations?
Risk Classification
ESL AnalysisFDA Class II via 510(k) pathway; IEC 62304 Class B (software safety classification); Post-§524B but no clear network capability detected
Learn how Parasoft automates IEC 62304 compliance testing →Cybersecurity Evidence
Powered by SBOMator™Based on AI-generated device summaries, not the full FDA submission. Items not detected here may still be present in the complete submission package.
Supply Chain BOM Evidence
Based on the CycloneDX / OWASP ECMA-424 xBOM ecosystem. Detection is from the AI-generated summary.
Analysis:
• Analysis is based on the AI-generated device summary, not the full FDA submission package. Click "Analyze Full FDA PDF" below to scan the actual submission document.
• Evidence of SBOM, ML-BOM, SaaSBOM, HBOM detected in AI-generated device summary.
• Postmarket cybersecurity monitoring plan not clearly documented in AI-generated device summary.
• Not detected in AI-generated device summary: CBOM, VDR/VEX, OBOM. These may be present in the full FDA submission.
Preparing a similar submission?
ESL takes complete ownership of your software evidence work - SBOM & CVE remediation with SBOMator™, static analysis & unit testing with Parasoft, traceability, and full IEC 62304 lifecycle support.
Assess Your Readiness →Or contact ESL directly